Code Alarm CR3 Manuale di Servizio Pagina 17

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 66
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 16
MIQ/CR3; DIQ/CR3 Settings
4 - 3
ba76032e01 01/2012
4.1.2 Limit indicator
With a limit indicator, a relay switches when a specified limiting value is
exceeded or undercut.
Limit indicators can be used in the following way:
Monitoring a limiting value using a relay:
when a limiting value (upper or lower limiting value) is exceeded or
undercut, a relay switches. The Open or Close relay actions are
possible in each case (see page 4-4).
Monitoring two limiting values using two relays:
If the upper limiting value is exceeded or undercut, a relay switches,
and if the lower limiting value is exceeded or undercut, another relay
switches. The Open or Close relay actions are possible in each case
(see page 4-4).
Note
If the simple monitoring function (Open, Close) with one or two relays
is not sufficient, use proportional output (see section 4.1.3).
Vedere la pagina 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 65 66

Commenti su questo manuale

Nessun commento